The Safety and Efficacy of Angioseal in Therapeutic Endovascular Interventions

  • M. Aksoy

      Affiliations

    • Department of General Surgery, Medical Faculty of Istanbul, Istanbul, Turkey
    • Corresponding Author InformationCorresponding author. Murat Aksoy, Istanbul Tip Fakultesi, Genel Cerrahi ABD Sekreterligi, 34390 Capa, Istanbul, Turkey
  • ,
  • J.-P. Becquemin

      Affiliations

    • Department of Vascular Surgery, Henri Mondor Hospital, Creteil, France
  • ,
  • P. Desgranges

      Affiliations

    • Department of Vascular Surgery, Henri Mondor Hospital, Creteil, France
  • ,
  • E. Allaire

      Affiliations

    • Department of Vascular Surgery, Henri Mondor Hospital, Creteil, France
  • ,
  • H. Kobeiter

      Affiliations

    • Department of Radiology, Henri Mondor Hospital, Creteil, France

Accepted 7 December 2005. published online 14 February 2006.

Article Outline

Abstract 

Objective

The aim of this study was to evaluate the safety and efficacy of the closure device Angioseal.

Material and methods

All consecutive patients, who underwent a therapeutic radiological intervention using the femoral artery approach from January 2001 to January 2005 in the Service of Vascular Surgery, Henri Mondor Hospital, Creteil, France, were prospectively included in the study. The efficacy of Angioseal was defined by the ability of the device to cover the puncture site and stop bleeding. The safety was defined by the rate of complications.

Results

A total of 79 Angioseal devices were employed on 77 patients. There were 62 (78.5%) males and 17 (21.5%) females. The mean age of the patients was 65.2±11.6 years (mean±SD). All Angioseal devices were deployed successfully. There were 62 (78.4%) 6F and 17 (21.6%) 8F sheaths employed during the procedures. There were two minor (2.5%) hematomas, one (1.2%) major hematoma and one (1.2%) pseudoaneurysm after the procedure. The mean time of discharge from the hospital was 2.1±1.8 (mean±SD) days. The patients were followed up for a mean of 9.0±9.3 (1–60 months) months.

Conclusions

Angioseal provides a safe and effective way of closing the femoral artery puncture site with acceptable morbidity rates.

Keywords: Closure devices, Angioseal, Vascular surgery, Radiology

 

Back to Article Outline

1. Introduction 

The traditional method of hemostasis following an interventional vascular procedure is manual compression. Puncture site complications include bleeding, hematoma, vessel occlusion and pseudoaneurysm formation. The prolonged duration of manual compression and a prolonged bed rest leads to patient discomfort.1, 2

A number of vascular closure devices are available.3, 4 Angioseal, which was introduced as a vascular closure device in Europe in 1994 operate by compressing the puncture site between an anchor and a collagen plug.5

This study aimed to evaluate the efficacy and safety of Angioseal in interventional radiological therapeutic interventions of the peripheral arterial system.

Back to Article Outline

2. Material and Methods 

All consecutive patients, who had undergone a therapeutic radiological intervention using the femoral artery approach from January 2001 to 2005 in the Service of Vascular and Endocrine Surgery, Henri Mondor Hospital, Creteil, France, were prospectively included in the study. The data regarding patient age, sex, demographics, and the use of anticoagulants and antiplatelet agents were recorded. The data concerning the procedural details such as the size of the vascular sheath employed, type of procedure performed, and indication for arteriography was noted.

A positive history of hypertension was defined as need for antihypertensive drugs or systolic blood pressure 160mm Hg or greater or diastolic blood pressure 95mm Hg or greater. A positive history of hypercholesterolemia was defined as need for cholesterol-lowering medication or serum cholesterol concentration 5.0mmol/L or greater. A positive history of diabetes mellitus was defined as need for glucose-lowering medication or (non-fasting) serum glucose concentration of 11.0mmol/L or greater. Currents smokers and former smokers in the last 10 years were categorized as smokers. A positive history of angina pectoris or previous cardiac intervention, ischemia on ECG and the need for a cardiac treatment were categorized as cardiac disease.

Patients underwent percutaneous approach using the standard Seldinger technique. The procedures were performed in the operating theatre or in the interventional radiology suite. Patients were prepped with betadine before the intervention. Heparin (0.5IU/kg) was administered intravenously before balloon dilatation or stent placement. The size of the Angioseal devices was decided according to the vascular sheath that was used. The puncture site was closed with 6-French Angioseal after the use of 5F and 6F sheaths whilst 8F Angioseal device was used for 7F sheaths. Patients were placed at bed rest for at least 30min after use of the 6F device and 90min following use of the 8F device according to the product representative recommendations.

The efficacy of Angioseal was defined by the ability of the device to cover the puncture site and stop bleeding. The safety was defined by the rate of complications. Bleeding was defined by a hematoma greater than 6cm and a decrease in hemoglobin of more than 1g, as well as those requiring transfusion or operation.

Back to Article Outline

3. Results 

A total of 79 Angioseal devices were employed on 77 patients during the study period. There were 62 (78.5%) males and 17 (21.5%) females. The mean age of the patients was 65.2±11.6 years (mean±SD). Of 79 patients, 9 (11.4%) had diabetes, 39 (49.3%) had hypertension, 17 (21.5%) had hyperlipidemia, 16 (20.2%) had coronary heart disease, and 22 (27.8%) were current smokers. There were 2 (2.5%) patients with bladder cancer, one patient (1.2%) with prostate cancer, one patient (1.2%) with Hodgkin's disease, one patient (1.2%) with larynx cancer, and another patient (1.2%) with cirrhosis Table 1. All patients were receiving antiplatelet treatment at the time of the intervention, whilst 2 (2.5%) of the patients were receiving anticoagulant treatment. Thirty-five (44.3%) patients were hospitalized for lower limb ischemia, whilst 27 (34.2%) and 17 (21.5%) patients were hospitalized for carotid stenosis and other causes, respectively.

Table 1. Patients' characteristics and risk factors
Hypertension39 (49.3%)
Current smokers22 (27.8%)
Hyperlipidemia17 (21.5%)
Coronary heart disease16 (20.2%)
Diabetes mellitus9 (11.4%)
Bladder cancer2 (2.5%)
Prostate cancer1 (1.2%)
Hodgkin's disease1 (1.2%)
Larynx cancer1 (1.2%)
Cirrhosis1 (1.2%)

The procedures included iliac and lower limb revascularization in 36 (45.5%) patients, cerebrovascular procedures in 28 (35.4%), and renal angioplasty and stenting in 10 (12.6%) patients. Two (2.5%) of the patients underwent two separate interventions during the procedure. The patients, who had undergone a procedure for lower limb revascularization, included 3 (3.7%) patients that received fibrinolysis for an occluded femoro-popliteal bypass. The details of the procedures are shown in Table 2.

Table 2. The details of the procedures, which have been performed during the study period, are shown
ProcedureFrequencyPercent
Carotid and iliac artery11.27
Iliac+superficial femoral artery11.27
Iliac artery2126.58
Lower limb revascularisation1620.25
Renal artery1012.66
Subclavian artery22.53
Superior mesenteric artery11.27
Vertebral artery11.27
Total79100

All Angioseal devices were deployed successfully. There were 62 (78.4%) 6F and 17 (21.6%) 8F sheaths employed during the procedures. There were two minor (2.5%) hematomas and one (1.2%) major hematoma after the procedure. The patient, who had a major hematoma required two units of blood transfusion but no surgical intervention. He also had a pseudoaneurysm of 5mm, which was demonstrated by computerized tomography scanning. The pseudoaneurysm resolved at 1 month. Two patients suffered from temporary renal insufficiency, which resolved with hydration. The mean time of discharge from the hospital was 2.1±1.8 (mean±SD) days (Table 2).

The patients were followed up for a mean of 9.0±9.3 (1–60 months) months. One patient, who had chronic renal insufficiency and a lower extremity ulcer died as a result of a septic shock 2 months after the intervention. There were no local complications during the follow-up period.

Back to Article Outline

4. Discussion 

Endovascular interventions have become a commonly accepted method of treatment in certain vascular diseases. Access site complications may lead to prolonged hospital stay and increased costs.1, 6 Hemostasis at the femoral artery access site is usually achieved by manual compression following angiographic and interventional cardiovascular procedures. Although manual compression is effective in achieving hemostasis in most cases, it requires a 15–20min sustained compression over the puncture site followed by prolonged rest.7 When manual compression is used, a number of standard protocols require a minimum of 4h of bed rest following diagnostic studies performed with 5F sheaths and a minimum of 6h of bed rest when larger sheaths are used for therapeutic interventions.7 Upper extremity fatigue and human resource considerations are other disadvantages of the method.

The drawbacks of manual compression have led to a search for a more comfortable methods of puncture site hemostasis for both the physician and the patient. The initial attempt was to reduce the sizes of catheters and guidewires. This led to an decreased rate of groin complications but the latest procedures in endovascular surgery required larger sheaths.8, 9 The frequent use of antiplatelet or anticoagulant agents created a tendency to bleeding in patients and an increase rate of access site complications.10, 11, 12 The first generation of vascular closure devices was not very welcome since these ‘unsophisticated’ devices were plagued with significant complication rates.13 The design has undergone some significant changes in order to achieve a safer and a more beneficial device. The Angioseal device consists of an anchor composed of a polylactide and polyglycolide polymer, a collagen plug, and suture contained within a special carrier system.5 When inserted, it achieves hemostasis by compressing the arterial puncture site between the anchor and the collagen plug. The Angioseal can consistently be deployed within 1min, and is completely bioabsorbable.

In our series, Angioseal was effective in achieving hemostasis in 98.7% of the patients. One patient developed a retroperitoneal hematoma, which required two units of blood transfusion and had a pseudoaneurysm of 5mm at the puncture site on the computerized tomography scan. However, the hematoma did not require further treatment than transfusion and the pseudoaneurysm resolved spontaneously at 1 month. The results of previous studies are in accordance to ours, revealing Angioseal as an effective and a safe method of femoral puncture site closure device following interventional vascular procedures. An efficacy of 95–100% exists in different reports.14 In a recent report, which included 188 patients and 144 procedures, all but three of Angioseal devices were deployed successfully.2 These results are satisfying because manual compression was effective in achieving hemostasis without any further complications in patients with failure of deployment. As well as its efficacy, its safety is promoted by these reports as complications such as infection, hemorrhage, and vessel occlusion have been uncommon, ranging from 0 to 1.9%.15, 16, 17 In the series of Abando et al.,2 which included 76 therapeutic interventions, there were no significant bleeding complications or infections.

Angioseal allowed early ambulation, rapid recovery time, and decreased discomfort in our patients. Hence, the patients may be more satisfied with safe and effective closure devices. The mean time of discharge of our patients was 2.1 days but this group also included patients with diabetic foot ulcers. Therefore, it included patients, who needed prolonged hospitalization for local treatment. However, in 1-day procedures, closure devices are supportive since patients can be ambulated 1h after the placement of a 6F device, and within 3h after placement of an 8F device.2

In conclusion, we believe that Angioseal provides a safe and effective way of closing the puncture site at the femoral artery following an interventional therapeutic vascular approach.

Back to Article Outline

References 

  1. Ricci MA, Trevisani GT, Pilcher DB. Vascular complications of cardiac catheterization. Am J Surg. 1994;167:375–378
  2. Abando A, Hood D, Weaver F, Katz S. The use of the Angioseal device for femoral artery closure. J Vasc Surg. 2004;40(2):287–290
  3. Semler HJ. Transfemoral catheterization: mechanical versus manual control of bleeding. Radiology. 1985;154:234–235
  4. Prayck JB, Wall TC, Longabaugh P, Tice FD, Hochrein J, Green C. A randomized trial of vascular hemostasis techniques to reduce femoral vascular complications after coronary intervention. Am J Cardiol. 1998;81:970–976
  5. Ward SR, Casale P, Raymond R, Kussmaul WG, Simpfendorfer C. Efficacy and safety of a hemostatic puncture closure device with early ambulation after coronary angiography. Angio-Seal Investigators. Am J Cardiol. 1998;81:569–572
  6. Eidt JF, Habibipour S, Saucedo JF, Mckee J, Southern F, Barone GW. Surgical complications from hemostatic puncture closure devices. Am J Surg. 1999;178:511–516
  7. Katz SG, Abando A. The use of closure devices. Surg Clin North Am. 2004;84(5):1267–1280[VI. Review]
  8. Kern MJ, Cohen M, Talley JD, Litvack F, Serota H, Aguirre F. Early ambulation after 5 French diagnostic cardiac catheterization: results of a multicenter trial. J Am Coll Cardiol. 1990;15:1475–1483
  9. Bogart DB, Bogart MA, Miller JT, Farrar MW, Barr WK, Montgomery MA. Femoralartery catheterization complications: a study of 503 consecutive patients. Catheter Cardiovasc Diagn. 1995;34:9–13
  10. EPILOG Investigators. Platelet glycoprotein IIB/IIIa receptor blockade and low-dose heparin during percutaneous coronary revascularization. N Engl J Med. 1997;336:1689–1696
  11. Aguirre FV, Topol EJ, Ferguson JJ, Anderson K, Blankenship JC, Heuser RR, et al. Bleeding complications with the chimeric antibody to platelet glycoprotein IIb/IIIa integrin in patients undergoing percutaneous coronary intervention. Circulation. 1995;91:2882–2890
  12. Standards of Practice Committee, Society of Cardiovascular and Interventional Radiology. Standard for diagnostic arteriography in adults. J Vasc Interv Radiol. 1993;4:385–395
  13. Heintzen MP, Straure BE. Peripheral arterial complications after heart catheterization. Herz. 1998;23:4–20
  14. Henry M, Amor M, Allaoui M, Tricoche O. A new access site management tool: the Angio-Seal hemostatic puncture closure device. J Endovasc Surg. 1995;2:289–296
  15. Shammas NW, Rajendran VR, Alldredge SG, Witcik WJ, Robken JA, Lewis JR. Randomized comparison of vasoseal and angioseal closure devices in patients undergoing coronary angiography and angioplasty. Catheter Cardiovasc Interven. 2002;55:421–425[Cremonesi A]
  16. Castriota F, Tarantino F, Troiani E, Ricci E, El Jamal B, et al. Femoral arterial hemostasis using the Angio-Seal system after coronary and vascular percutaneous angioplasty and stenting. J Invasive Cardiol. 1998;10:464–469
  17. Bown MJ, Blanshard KS, Cutress ML, Sayers RD. Off-license use of Angio-Seal arterial puncture closure device. Eur J Vasc Endovasc. 2002;24:372–373

PII: S1078-5884(06)00006-2

doi:10.1016/j.ejvs.2005.12.014

European Journal of Vascular & Endovascular Surgery
Volume 32, Issue 1 , Pages 90-93, July 2006