Early Results from the ENGAGE Registry: Real-world Performance of the Endurant Stent Graft for Endovascular AAA Repair in 1262 Patients

Open ArchivePublished:July 26, 2012DOI:https://doi.org/10.1016/j.ejvs.2012.07.005

      Abstract

      Objective

      The ENGAGE registry was undertaken to examine the real-world outcome after endovascular abdominal aortic aneurysm (AAA) repair (EVAR) with the Endurant Stent Graft in a large, contemporary, global series of patients.

      Methods

      From March 2009 to April 2011, 1262 AAA patients (89.6% men; mean age 73.1 years, range 43–93 years) were enrolled from 79 sites in 30 countries and treated with Endurant. Results are described following the reporting standards for EVAR. Follow-up data were tabulated for all 1262 patients at a 30-day follow-up and for the first 500 patients at a 1-year follow-up.

      Results

      Intra-operative technical success was achieved in 99.0% of cases. Within 30 days, adverse events were reported in 3.9% of patients, including a 1.3% mortality rate. Type-I or –III endoleaks were identified in 1.5% of cases. Estimated overall survival, aneurysm-related survival and freedom from secondary interventions at 1 year were 91.6%, 98.6% and 95.1%, respectively. At 1 year, aneurysm size increased ≥5 mm in 2.8% and decreased ≥5 mm in 41.3% of cases.

      Conclusion

      Early results from this real world, global experience are promising and indicate that endovascular AAA repair with the Endurant Stent Graft is safe and effective across different geographies and standards of practice. Longer-term follow-up is necessary to assess durability of these results.

      Keywords

      • The technology of endovascular aneurysm repair (EVAR) for abdominal aortic aneurysms (AAAs) is a dynamic, ever-changing endeavour. The challenge is to decrease complications and re-interventions while safely treating more complex anatomy. As improved devices become available and operators become more proficient with endovascular techniques, EVAR outcomes change. In order to make balanced judgement about the management of AAAs, it important to augment the knowledge base about EVAR in contemporary practice with the latest generation devices. The Endurant Stent Graft Natural Selection Global Postmarket Registry (ENGAGE) has been designed to closely monitor the real-world performance of the Endurant Stent Graft System. Unprecedented in size, scope and geographic representation, it represents a combined experience of 79 sites worldwide.
      Abdominal aortic aneurysms (AAAs) are a significant health challenge with an estimated incidence of 20–40 cases per 100 000 population per year. Patients present with varying levels of risk due to aneurysm size, age and concurrent co-morbidities.
      • Lederle F.A.
      • Johnson G.R.
      • Wilson S.E.
      • Chute E.P.
      • Littooy F.N.
      • Bandyk D.
      • et al.
      Prevalence and associations of abdominal aortic aneurysm detected through screening. Aneurysm Detection and Management (ADAM) Veterans Affairs Cooperative Study Group.
      Since Parodi
      • Parodi J.C.
      • Palmaz J.C.
      • Barone H.D.
      Transfemoral intraluminal graft implantation for abdominal aortic aneurysms.
      and Volodos
      • Volodos N.L.
      • Karpovich I.P.
      • Troyan V.I.
      • Kalashnikova Yu V.
      • Shekhanin V.E.
      • Ternyuk N.E.
      • et al.
      Clinical experience of the use of self-fixing synthetic prostheses for remote endoprosthetics of the thoracic and the abdominal aorta and iliac arteries through the femoral artery and as intraoperative endoprosthesis for aorta reconstruction.
      published the first transfemoral intraluminal graft implantations for AAA in 1991, the practice of endovascular aneurysm repair (EVAR) has continually improved. Today, this approach is a generally accepted alternative to conventional open surgical repair due to reductions in perioperative mortality and morbidity, blood loss, use of the intensive care unit (ICU) and length of hospital stay.
      • Chaikof E.L.
      • Brewster D.C.
      • Dalman R.L.
      • Makaroun M.S.
      • Illig K.A.
      • Sicard G.A.
      • et al.
      The care of patients with an abdominal aortic aneurysm: the Society for Vascular Surgery practice guidelines.
      Several randomised trials have confirmed these benefits,
      • EVAR-1 trial investigators
      Endovascular aneurysm repair versus open repair in patients with abdominal aortic aneurysm (EVAR trial 1): randomised controlled trial.
      • Prinssen M.
      • Verhoeven E.L.
      • Buth J.
      • Cuypers P.W.
      • van Sambeek M.R.
      • Balm R.
      • et al.
      A randomized trial comparing conventional and endovascular repair of abdominal aortic aneurysms.
      • Lederle F.A.
      • Freischlag J.A.
      • Kyriakides T.C.
      • Padberg Jr., F.T.
      • Matsumura J.S.
      • Kohler T.R.
      • et al.
      Outcomes following endovascular vs open repair of abdominal aortic aneurysm: a randomized trial.
      and one trial confirmed these benefits except for perioperative mortality and major adverse events.
      • Becquemin J.P.
      • Pillet J.C.
      • Lescalie F.
      • Sapoval M.
      • Goueffic Y.
      • Lermusiaux P.
      • et al.
      A randomized controlled trial of endovascular aneurysm repair versus open surgery for abdominal aortic aneurysms in low- to moderate-risk patients.
      The success of EVAR, however, is dependent upon patient-specific factors, including morphology and dimensions of the aneurysm. Severely angulated or short infrarenal aortic necks and small, tortuous or calcified iliac arteries are related to adverse EVAR outcomes,
      • Choke E.
      • Munneke G.
      • Morgan R.
      • Belli A.M.
      • Loftus I.
      • McFarland R.
      • et al.
      Outcomes of endovascular abdominal aortic aneurysm repair in patients with hostile neck anatomy.
      and thus guidelines for commercially available stent grafts indicate use within a specific range of anatomy. A substantial portion of AAA patients fall outside these generally accepted inclusion criteria,
      • Carpenter J.P.
      • Baum R.A.
      • Barker C.F.
      • Golden M.A.
      • Mitchell M.E.
      • Velazquez O.C.
      • et al.
      Impact of exclusion criteria on patient selection for endovascular abdominal aortic aneurysm repair.
      and their advanced disease state or major co-morbidities present a high risk for open surgical repair. Therefore, there is a need for improved stent grafts and endovascular techniques to decrease complications and increase eligibility.
      Medtronic Endovascular (Santa Rosa, CA, USA) designed its latest generation product, the Endurant Stent Graft System, to address the limitations of previous stent graft designs. A small amplitude M-shaped proximal stent was designed to improve sealing at the proximal neck while potentially allowing for greater sizing flexibility. Radial strength was also improved while allowing a lower-profile delivery system. The Endurant Stent Graft System received CE mark approval in July 2008 and Food and Drug Administration (FDA) approval in December 2010. After a safety assessment trial conducted in Europe,
      • Rouwet E.V.
      • Torsello G.
      • de Vries J.P.
      • Cuypers P.
      • van Herwaarden J.A.
      • Eckstein H.H.
      • et al.
      Final results of the prospective European trial of the Endurant stent graft for endovascular abdominal aortic aneurysm repair.
      the Endurant Stent Graft Natural Selection Global Postmarket Registry (ENGAGE) was undertaken to quantify the performance of this recently released endograft within the context of contemporary, real-world use. Herein, we report the perioperative and 1-year results of this global, multicentre, prospective 1262-patient study of the Endurant Stent Graft System.

      Methods

      Patient population

      Between March 2009 and April 2011, eligible patients from 79 sites in 30 countries (Appendix) were enrolled in the ENGAGE registry. The study protocol strongly encouraged consecutive enrolment of at least five patients in a row to minimise selection bias. Ruptured AAAs were not considered for enrolment into ENGAGE. Prior to the index procedure, computed tomography angiography (CTA) imaging of the abdomen and pelvis was undertaken to determine the morphological eligibility for elective endovascular treatment with an Endurant Stent Graft. To reflect the real-world clinical practice, eligibility criteria for participation were kept comprehensive. Although the study design included that individual morphological variables (proximal neck diameter and length, infrarenal and suprarenal angulation and distal iliac fixation diameter and length) were consistent with Endurant’s Instructions for Use (IFU) (Table 1), enrolment of patients who fell outside the IFU guidance was accepted. Patients considered unlikely to adhere to the follow-up regimen and patients with concurrent trial participation were excluded from study enrolment. A signed consent for authorisation of data release was required. The trial was conducted according to the Declaration of Helsinki and the International Conference on Harmonization (ICH) Good Clinical Practice (GCP) guidelines, and approved by local medical ethics committees.
      Table 1Endurant® indications for use.
      • One of the following criteria:
      • ■ Infrarenal neck length ≥10 mm with non-significant calcification, and/or non-significant thrombus in combination with ≤45° suprarenal angulation and ≤60° infrarenal angulation.
      • ■ Infrarenal neck length ≥15 mm with non-significant calcification and/or non-significant thrombus in combination with ≤60° suprarenal angulation and ≤75° infrarenal angulation.
      • And all of the following criteria:
      • ■ Adequate iliac/femoral access
      • ■ Proximal AAA neck diameter ≥19 mm and ≤32 mm
      • ■ Distal iliac fixation site diameter ≥8 mm and ≤25 mm
      • ■ Distal non-aneurysmal iliac fixation length ≥15 mm bilaterally

      Study procedure

      Technical specifications of the Endurant Stent Graft System have been published previously.
      • Rouwet E.V.
      • Torsello G.
      • de Vries J.P.
      • Cuypers P.
      • van Herwaarden J.A.
      • Eckstein H.H.
      • et al.
      Final results of the prospective European trial of the Endurant stent graft for endovascular abdominal aortic aneurysm repair.
      • Bockler D.
      • Fitridge R.
      • Wolf Y.
      • Hayes P.
      • Silveira P.G.
      • Numan F.
      • et al.
      Rationale and design of the Endurant Stent Graft Natural Selection Global Postmarket Registry (ENGAGE): interim analysis at 30 days of the first 180 patients enrolled.
      Preoperatively, each patient had a customised plan made with respect to stent diameters and length, dependent on their aortic and iliac dimensions. For optimal sealing, it was advised to oversize the stent diameter by approximately 20% based on inner vessel diameter. All endovascular procedures were performed under fluoroscopic control. Local, regional or general anaesthetics, antibiotics and heparin were administered according to each site’s standard regimen. Arterial access was conducted by operator’s preference. If necessary, coil embolisation of the hypogastric or inferior mesenteric artery, or other adjunctive procedures were performed and documented before or during the implant procedure. A completion angiogram was performed to document the status after stent graft implantation.

      Imaging and follow-up

      Follow-up was planned according to standard practice at each clinical site, with the exception of the requirement for 30-day and 1-year imaging studies No specific tests or procedures that fell outside a site’s standard regimen for AAA follow-up were required. Diagnostic images were analysed at both time points for technical outcomes and AAA changes. The presence of endoleak was classified by type and recorded as well as changes in aneurysm size.

      ‘End’ points and definitions

      The ENGAGE registry was designed to assess effectiveness of the Endurant Stent Graft at 12 months post-implantation, with follow-up extended to 5 years. This manuscript describes the initial procedural data and outcomes, the technical observations, adverse events and major adverse events (MAEs) within 30 days and 12 months, along with stent graft migration and aneurysm expansion (>5 mm) between 30 days and 12 months.
      The primary effectiveness ‘end’ point was the initial procedural success, a composite of technical success and clinical success at the time of the index procedure. Technical success was defined as successful delivery and deployment of the Endurant Stent Graft in the planned position without unintentional coverage of one or both internal iliac arteries or visceral aortic branches and with successful removal of the delivery system. In the case of primary conversion, technical success failed. Initial clinical success was defined as technical success without intra-operative death or presence of a type-I/III endoleak at the conclusion of the index procedure.
      The secondary ‘end’ points consisted of technical observations, adverse events and MAE. Technical observations included: stent graft kinking; stent graft wire form fracture; suprarenal bare stent fracture or detachment from fabric; occlusion (defined as 100% obstruction); stenosis (defined as partial obstruction); and presence of an endoleak. Secondary conversion to open repair, aneurysm rupture and secondary interventions were stated as adverse events. All-cause mortality, bowel ischaemia, myocardial infarction, paraplegia, renal failure (requiring dialysis or elevated serum creatinine two times baseline value), respiratory failure (need for >24 h mechanical ventilation postoperatively or reintubation for any reason), stroke and procedural blood loss ≥1000 cc were noted as MAEs. All deaths within 30 days postoperative were judged to be aneurysm-related. All-cause mortality, aneurysm-related mortality and secondary procedures were separately assessed as 1-year Kaplan–Meier estimates.

      Data management, quality control and statistical analysis

      Data collected on each patient were recorded on a web-based electronic case report form (Veracity™ Clinical Asset Management, MERGE Healthcare, Chicago, IL, USA) to ensure reliable data collection, data management, secure authentication and traceability. Data were entered by, or under supervision of, sites’ principal investigators. Data management and biostatistical analysis were performed by the Medtronic Biostatistics & Data-Management Department (Santa Rosa, CA, USA). It reviewed 100% of data to detect missing or inconsistent data to generate queries to the investigators for resolution. In addition to this, Medtronic Bakken Research Centre BV (Maastricht, the Netherlands) randomly monitored over 40% of patients’ source documentation against the data entered. They also performed a verification of all 1262 patient informed consents and essential study documents at each site.
      All variables are reported descriptively with no hypothesis testing. For categorical variables, frequency and percentage were calculated. For continuous variables, mean, standard deviation, minimum and maximum were calculated. A per protocol analysis was performed for the technical observations. All other variables were evaluated on an intention-to-treat basis.

      Results

      From March 2009 until April 2011, 1266 patients from 79 sites in 30 countries were initially recorded in the database. Four patients however, were excluded from the study for the following reasons; one patient refused immediate treatment, but underwent emergency EVAR for a ruptured AAA 3 months later; one patient underwent open surgical treatment instead, because of unsuitable anatomy on preoperative imaging; one patient was operated on at a non-participating site; and with one patient, the informed consent form was missing. Data for the remaining 1262 patients comprise the basis of this report. At time of writing, a sub-cohort of the first 500 operated patients (39.6%) had the opportunity to complete 12-month follow-up. Their data form the basis of the 1-year outcome analysis.

      Baseline characteristics

      Patients’ demographics and risk factors (Table 2) were typical of an AAA population. They were predominantly male, elderly and American Society of Anaesthesiologists (ASA) class II or III, with a high prevalence of cardiovascular risk factors and co-morbidities. The primary indication for EVAR was an AAA diameter >5 cm in 88.1% of cases and 83.9% of patients were asymptomatic. Table 3 describes baseline aneurysm characteristics. The mean maximum AAA diameter was 60.3 ± 11.7 mm. The proximal aortic neck had a mean diameter of 23.7 ± 3.6 mm, with a non-aneurysmal length of 27.0 ± 12.4 mm (27 (2.2%) patients with neck length <10 mm) and a mean infrarenal neck angulation of 30.3 ± 23.8°. Notably, 226 patients (17.9%) were implanted outside the IFU criteria (Table 4).
      Table 2Patient demographics and risk factors (ITT analysis).
      VariableN = 1262
      Denominator differs when there are missing values.
      Age (years) Mean ± SD (range)73.1 ± 8.1(43–93)
      Gender
       Male89.6%(1131/1262)
       Female10.4%(131/1262)
      ASA
      American Society of Anesthaesiologists.
      Classification
       Class I6.1%(77/1261)
       Class II41.8%(527/1261)
       Class III41.5%(523/1261)
       Class IV10.6%(134/1261)
      Symptoms
       A-symptomatic AAA
      Abdominal aortic aneurysm.
      83.9%(1059/1262)
       Symptomatic AAA
      Abdominal aortic aneurysm.
      16.1%(203/1262)
      Indication by AAA
      Abdominal aortic aneurysm.
      diameter
       >5 cm88.1%(1112/1262)
       4–5 cm (≥0.5 cm increase in last 6 months)6.3%(79/1262)
       ≥1.5x reference infrarenal aorta2.9%(42/1262)
       Other2.3%(29/1262)
      Risk factors
       Tobacco use49.3%(607/1231)
       Hypertension75.4%(939/1245)
       Hyperlipidaemia60.4%(718/1188)
       Diabetes19.0%(236/1244)
       Cancer20.5%(254/1241)
       Cardiac disease
        Myocardial infarction (MI)26.3%(318/1210)
        Arrhythmia16.0%(198/1234)
        Coronary artery disease (CAD)34.6%(422/1218)
        Cardiac revascularisation27.1%(337/1244)
       Pulmonary disease25.1%(311/1241)
       Renal insufficiency15.3%(191/1251)
       Cerebrovascular disease
        Transient ischaemic attack (TIA)4.9%(61/1249)
        Cerebral vascular accident (CVA)5.3%(67/1255)
      Gastro-intestinal complications19.5%(246/1261)
      a Denominator differs when there are missing values.
      b American Society of Anesthaesiologists.
      c Abdominal aortic aneurysm.
      Table 3Baseline aneurysm characteristics (ITT analysis).
      VariableMean ± SDN = 1262 (range)
      Maximum AAA
      Abdominal aortic aneurysm.
      diameter (mm)
      60.3 ± 11.7(30–118)
      Proximal neck diameter (mm)23.7 ± 3.6(15–45
      Upper range of 45 mm: Tapered aortic neck, with a proximal diameter of 45 mm, a diameter of 22 mm immediately above the aneurysm and a 15 mm non-aneurysmal neck length.
      )
      Proximal non-aneurysmal neck length (mm)27.0 ± 12.4(0
      Lower range of 0 mm: Case performed with a “chimney technique”.
      – 80)
      Distal iliac fixation site diameter (mm)
       Right14.1 ± 3.6(2–29)
       Left13.8 ± 3.5(6–30)
      Infrarenal neck angle (°)30.3 ± 23.8(0–130)
       ≤60°89.8%
       60–75°6.0%
      Suprarenal neck angle (°)18.8 ± 18.5(0–120)
       ≤45°92.5%
       45–60°4.5%
      a Abdominal aortic aneurysm.
      b Upper range of 45 mm: Tapered aortic neck, with a proximal diameter of 45 mm, a diameter of 22 mm immediately above the aneurysm and a 15 mm non-aneurysmal neck length.
      c Lower range of 0 mm: Case performed with a “chimney technique”.
      Table 4Patients implanted outside IFU
      Instructions for use.
      (ITT analysis).
      VariableN = 1262
      Total implanted outside of IFU
      Instructions for use.
      guidance
      17.9%(226/1262)
       Non-primary indications2.3%(29/1262)
       Proximal neck diameter < 19 mm or >32 mm5.7%(72/1255)
       Proximal neck length < 10 mm2.2%(27/1248)
       Proximal neck length ≥ 10 mm and <15 mm
      In combination with suprarenal angle > 45° or infrarenal angle > 60°.
      2.1%(26/1248)
       Angulation suprarenal > 60° or infrarenal > 75°4.7%(59/1248)
       Distal iliac fixation site diameter <8 mm1.5%(19/1262)
       Distal iliac fixation site diameter >25 mm0.6%(8/1262)
      a Instructions for use.
      b In combination with suprarenal angle > 45° or infrarenal angle > 60°.

      Intra-operative outcome

      The procedure was performed under general, regional and local anaesthesia in, respectively, 62.3%, 26.2% and 11.5% of cases. The mean procedural duration was 99.5 ± 45.0 min (range 20–387 min), with a mean total fluoroscopic time of 20.5 ± 12.4 min (range 0–92 min) and a mean contrast volume used of 130.7 ± 70.6 ml (range 0–400 ml). Mean intra-operative blood loss was 208.4 ± 220.1 ml (range 0–2700 ml). ICU admission rate was 34.2%, and 6.0% were admitted longer than 24 h. The mean hospital stay from the date of the initial procedure was 4.83 ± 5.07 days (median 4.00; range 0.5–68 days).
      Technical success was achieved in 1250 patients (99.0%). The Endurant was delivered to the planned location and successfully deployed in 99.4% of patients (Table 5). Attempts at endovascular repair were discontinued in six (0.5%) patients: four cases with access problems due to tortuous or stenotic iliac arteries, two of whom were eventually converted to open repair after 2 and 49 days respectively. One patient required immediate conversion after unintentional coverage of both renal arteries, when the suprarenal stent was released before accurate placement of the covered stent. In another case, one of the proximal struts got stuck in the tip-capture portion of the delivery system, resulting in the inability to remove the device and necessitating immediate conversion to open surgery. There were no intra-operative deaths.
      Table 5Initial procedural data and evaluation (ITT & PP analysis).
      Placement procedure (ITT)N = 1262
      Pre-implant adjunctive procedure performed7.6%(96/1262)
       Coil embolization IIA
      Internal iliac arteries.
      4.3%(54/1262)
       Coil embolization IMA
      Inferior mesenteric arteries.
      1.1%(14/1262)
       Other2.4%(30/1262)
      Associated procedures performed during procedure
       Coil embolization IIA5.1%(64/1262)
       Coil embolization IMA0.6%(7/1262)
       Other8.7%(110/1262)
       None86.5%(1091/1262)
      Additional device used during implant procedure
       Balloon catheter76.5%(965/1262)
        Unplanned6.0%(76/1262)
       Stent (other than Endurant)6.2%(78/1262)
        Unplanned3.5%(44/1262)
       Other6.8%(86/126)
      Placement of proximal end of device
       With suprarenal stent crossing both renal arterie75.7%(955/1262)
       With suprarenal stent crossing one renal artery6.0%(76/1262)
       Below both renal arteries17.5%(221/1262)
      Placement of distal end of device
       Right limb distal to the IIA8.8%(111/1262)
       Left Limb distal to the IIA8.3%(105/1262)
      Endurant stent-graft implanted into a patient99.5%(1256/1262)
       2 components implanted43.0%(540/1256)
       3 components implanted27.5%(345/1256)
       4 components implanted21.9%(275/1256)
       ≥5 components implanted6.6%(83/1256)
      Primary effectiveness endpoints (ITT)N = 1262
      Technical success99.0%(1250/1262)
       Endurant stent-graft successfully delivered99.4%(1255/1262)
       Endurant stent-graft successfully deployed99.4%(1255/1262)
       No unintended coverage of IIA or any VABs
      Visceral aortic branches.
      99.5%(1253/1259)
       Endurant delivery system successfully removed99.9%(1261/1262)
      Intra-operative clinical success97.6%(1232/1262)
       Technical success99.0%(1250/1262)
       Freedom from intra-operative death100.0%(1262/1262)
       Freedom from type I/III endoleak98.6%(1238/1256)
      Initial technical observations (PP)N = 1256
      Stent graft malfunctions
       Kinking1.0%(13/1253)
       Twisting0.5%(6/1252)
       Wire form fracture0.0%(0/1253)
       Suprarenal bare stent fracture0.0%(0/1253)
       Other malfunctions0.3%(4/1256)
       Endoleak (Uncorrected)
      Uncorrected: Detected, but chosen not to treat within initial procedure.
      16.0%(201/1256)
        Type I1.1%(14/1256)
        Type II12.4 %(156/1256)
        Type III0.3%(4/1256)
        Type IV1.8%(22/1256)
        Undetermined0.7 %(9/1256)
      a Internal iliac arteries.
      b Inferior mesenteric arteries.
      c Visceral aortic branches.
      d Uncorrected: Detected, but chosen not to treat within initial procedure.
      Adjunctive procedures were performed before or during implant procedures in 96 (7.6%) and 171 (13.5%) cases, respectively (Table 5). Primarily coil embolisation of the internal iliac artery or inferior mesenteric artery. Unplanned additional stents (other than Endurant) were placed in 44 (3.5%) patients, mainly to resolve intra-operative type-I/III endoleaks. The final completion angiogram revealed type-I and type-III endoleaks in 14 (1.1%) and four (0.3%) cases, respectively, predominantly in patients with regular neck configurations. Among patients with type-I endoleaks, one patient underwent secondary placement of a proximal cuff. Another patient died of a myocardial infarction before 30-day imaging was performed. In all remaining cases, the endoleaks were absent at 30-day CTA and no secondary interventions were performed.

      Perioperative outcome

      One-month imaging was performed in 91.6% of the 1256 implanted patients (Table 6). One or more graft-related problems, including endoleaks, were reported in 191 (16.6%) of these patients. Occlusion of one of the iliac limbs was reported in 23 patients (2.0%) and graft stenosis was detected in 16 (1.4%). Stent graft kinking occurred in 20 (1.7%) cases. One report of occlusion of a (right) renal artery occurred. This did not result in major clinical event or re-intervention. There were no stent fractures or stent graft migrations reported in any patient through 30 days.
      Table 6Technical performance at 30-days (PP analysis).
      VariableN = 1151
      Only implanted patients with 1-month imaging study included in the analyses.
      One or more technical observations16.6%(191/1151)
      Stent graft kinking1.7%(20/1151)
      Stent graft twisting0.1%(1/1151)
      Stent graft wire form fracture0.0%(0/1151)
      Suprarenal bare stent fracture0.0%(0/1151)
      Suprarenal bare stent detachment from fabric0.1%(1/1151)
      Stent graft occlusion2.0%(23/1151)
      Stent graft stenosis1.4%(16/1151)
      Stent graft migration0.0%(0/1151)
      Endoleak12.0%(138/1151)
       Type I1.4%(16/1151)
       Type II9.9%(114/1151)
       Type III0.2%(2/1151)
       Type IV0.1%(1/1151)
       Undetermined0.6%(7/1151)
       Type I and/or III1.5%(17/1151)
      Other technical observation0.6%(7/1151)
      a Only implanted patients with 1-month imaging study included in the analyses.
      Endoleaks were present in 138 (12.0%) patients at 30 days, of which in seven cases the type could not be specified. type-I and/or -III endoleaks were identified in 17 (1.5%) patients, including one patient with both types present. One of the type-I endoleaks had been observed at the time of operation and persisted to day 32, when it was successfully treated. The remaining endoleaks were newly diagnosed on the 30-day imaging study. Among these, eight type-I endoleaks were treated with an extension or remodelling of the graft, four spontaneously resolved without intervention and two had not yet reached the 12-month follow-up visit. The patient with both a type-I and -III endoleak present had persistent endoleaks at the 12-month images, and was scheduled for re-intervention thereafter. The other type-III endoleak was resolved on follow-up images.
      The overall 30-day mortality rate was 1.3. One or more major adverse events within 30 days, including all-cause mortality, were reported in 3.9% of patients (Table 7); bowel ischaemia was recorded in three patients (0.2%), myocardial infarction in 14 patients (1.1%), renal failure in four patients (0.3%), stroke in two patients (0.2%) and blood loss ≥1000 cc was recorded in 18 (1.4%) patients. No patient developed paraplegia or respiratory failure.
      Table 7Patient outcome within 30-days (ITT analysis).
      VariableN = 1262
      One or more major adverse events (MAE)3.9%(49/1262)
       All-cause mortality1.3%(16/1262)
       Bowel ischemia0.2%(3/1262)
       Myocardial infarction1.1%(14/1262)
       Paraplegia0.0%(0/1262)
       Renal failure0.3%(4/1262)
       Respiratory failure0.0%(0/1262)
       Stroke0.2%(2/1262)
       Procedural blood loss ≥1000 cc1.4%(18/1262)
      Conversion to open surgery0.2%(3/1262)
      Secondary surgical procedure1.5%(19/1262)
       Endovascular for occlusion, stenosis or kinking0.6%(8/1262)
       Endovascular to correct Type I/III endoleak0.3%(4/1262)
       Open bypass procedure0.5%(6/1262)
       Other0.1%(1/1262)
      Aneurysm rupture0.0%(0/1262)
      Within the first month, three (0.2%) cases were converted to open surgery, two as emergencies and one electively, as mentioned earlier. Secondary interventions were required in 19 (1.5%) cases, within the first 30 days of follow-up; eight (0.6%) cases required endovascular intervention for graft occlusion, stenosis or kinking; one case was for contralateral leg implant; six (0.5%) cases required a bypass procedure; and four (0.3%) cases required endovascular intervention to correct a type-I endoleak.

      One-year outcome

      A sub-cohort of 500 patients was followed for at least 12 months, with compliance to follow-up of 98.6%. Baseline characteristics and perioperative outcome measures of this sub-cohort were comparable to those of the total study population. Therefore, further analysis on this sub-cohort was considered to be valuable.
      The 1-year Kaplan–Meier estimate for overall survival was 91.6 ± 1.4%. The 1-year estimate for aneurysm-related survival was 98.8 ± 0.5%, with no device-related deaths (Fig. 1). One or more MAEs, including all-cause mortality, were reported in 11.2% of patients within 1-year follow-up; including five (1.0%) cases of renal failure, nine (1.8%) myocardial infarctions, two (0.4%) cases of stroke, two (0.4%) case of bowel ischaemia and one (0.2%) case of respiratory failure.
      Figure thumbnail gr1
      Figure 1Kaplan–Meier estimates for all-cause mortality & AAA-related mortality.
      The Kaplan–Meier estimate for 1-year secondary intervention-free survival was 95.1 ± 1.1% (Fig. 2). Within the first year after implantation, secondary interventions were required in 23 (4.6%) patients. Endovascular procedures were performed in 10 patients to resolve graft occlusion, stenosis or kinking; in six patients to correct a type-I/III endoleak; and in three patients to resolve a persistent type-II endoleak. Five patients underwent a by-pass procedure for an occluded iliac limb. Notably, stent graft migration or loss of device integrity was never observed within the first year after implantation. In addition, there were no reports of aneurysm rupture. At 1-year, aneurysm size increased by ≥5 mm in 2.8% of cases, was stable in 55.9% of cases and decreased by ≥5 mm in 41.3% of cases.
      Figure thumbnail gr2
      Figure 2Kaplan–Meier estimates for secondary procedures.

      Discussion

      The technology of EVAR for AAAs is a dynamic, ever-changing endeavour. The challenge is to decrease complications and re-interventions while safely treating more complex anatomy, especially for those cases unfit for open repair. As improved devices with wider inclusion criteria become available and operators become more proficient with endovascular techniques, the proportion of patients suitable for EVAR increases.
      • van Marrewijk C.J.
      • Leurs L.J.
      • Vallabhaneni S.R.
      • Harris P.L.
      • Buth J.
      • Laheij R.J.
      Risk-adjusted outcome analysis of endovascular abdominal aortic aneurysm repair in a large population: how do stent-grafts compare?.
      However, overall outcomes may not improve if broadening the application rate results in poorer outcomes despite improvements in operator skill and device design. This ever-evolving trade-off implies that results from older series using previous generations of devices may not reflect the current status of EVAR. As a consequence, results of the performance of latest generation devices in contemporary, real-world settings are important. High-quality, well-designed observational studies are increasingly believed to provide complementary evidence to randomised controlled trials (RCTs).
      • Ligthelm R.J.
      • Borzi V.
      • Gumprecht J.
      • Kawamori R.
      • Wenying Y.
      • Valensi P.
      Importance of observational studies in clinical practice.
      Therefore, they are essential in making a balanced judgement about the management of AAAs.
      The ENGAGE registry was undertaken to quantify the performance of a recently released endograft within the context of contemporary, real-world use. This prospective observational study represents the collective experience of 79 centres in 30 countries across five different continents with the Endurant Stent Graft. As eligibility was for the most part, left to the discretion of the investigator, the outcomes in the 1262 AAA patient series are expected to have high external validity. Follow-up protocols were kept as close as possible to standard site regimens, to keep treatment to real world practice. In order to be able to produce meaningful analysis, a large quantity of data was recorded. To guarantee high quality and completeness of data, efforts were made to achieve onsite quality control and continual monitoring of reported data, which is evident in the high level of compliance to follow-up in this registry.
      The necessity for secondary interventions is considered the Achilles’ heel of EVAR.
      • De Bruin J.L.
      • Baas A.F.
      • Buth J.
      • Prinssen M.
      • Verhoeven E.L.
      • Cuypers P.W.
      • et al.
      Long-term outcome of open or endovascular repair of abdominal aortic aneurysm.
      • Peterson B.G.
      • Matsumura J.S.
      • Brewster D.C.
      • Makaroun M.S.
      Five-year report of a multicenter controlled clinical trial of open versus endovascular treatment of abdominal aortic aneurysms.
      Secondary interventions were the main reason that EVAR was not considered cost-effective at long-term follow-up in the DREAM (Dutch Randomised Endovascular Aneurysm Management) trial.
      • Prinssen M.
      • Buskens E.
      • de Jong S.E.
      • Buth J.
      • Mackaay A.J.
      • van Sambeek M.R.
      • et al.
      Cost-effectiveness of conventional and endovascular repair of abdominal aortic aneurysms: results of a randomized trial.
      A re-intervention rate of 4.6% is comparable to re-intervention rates of earlier reports and the recent OVER (Open Versus Endovascular Repair) trial, and compares favourably to the landmark studies EVAR −1 (Endovascular Repair versus Open Repair) and DREAM.
      • EVAR-1 trial investigators
      Endovascular aneurysm repair versus open repair in patients with abdominal aortic aneurysm (EVAR trial 1): randomised controlled trial.
      • Prinssen M.
      • Verhoeven E.L.
      • Buth J.
      • Cuypers P.W.
      • van Sambeek M.R.
      • Balm R.
      • et al.
      A randomized trial comparing conventional and endovascular repair of abdominal aortic aneurysms.
      • Lederle F.A.
      • Freischlag J.A.
      • Kyriakides T.C.
      • Padberg Jr., F.T.
      • Matsumura J.S.
      • Kohler T.R.
      • et al.
      Outcomes following endovascular vs open repair of abdominal aortic aneurysm: a randomized trial.
      • Kvinlaug K.E.
      • Lawlor D.K.
      • Forbes T.L.
      • Willoughby R.
      • MacKenzie K.S.
      • DeRose G.
      • et al.
      Early results from a Canadian multicenter prospective registry of the Endurant stent graft for endovascular treatment of abdominal aortic aneurysms.
      • Torsello G.
      • Troisi N.
      • Tessarek J.
      • Torsello G.F.
      • Dorigo W.
      • Pulli R.
      • et al.
      Endovascular aortic aneurysm repair with the Endurant stent-graft: early and 1-year results from a European multicenter experience.
      • van Keulen J.W.
      • de Vries J.P.
      • Dekker H.
      • Goncalves F.B.
      • Moll F.L.
      • Verhagen H.J.
      • et al.
      One-year multicenter results of 100 abdominal aortic aneurysm patients treated with the Endurant stent graft.
      The type II endoleak rate of 9.9%, which is remarkably lower than in these older studies, might have contributed to the lower rate of secondary procedures. Comparison of re-intervention rates at 1-year with other studies is hampered by several factors; in particular, the treatment of type-II endoleaks has changed over time. On the other hand, in the ENGAGE Registry, 17.9% of patients were treated outside IFU, primarily due to complex anatomy; therefore, it should also be taken into consideration that the eligibility criteria for ENGAGE were less strict than for DREAM and OVER.
      • Prinssen M.
      • Verhoeven E.L.
      • Buth J.
      • Cuypers P.W.
      • van Sambeek M.R.
      • Balm R.
      • et al.
      A randomized trial comparing conventional and endovascular repair of abdominal aortic aneurysms.
      • Lederle F.A.
      • Freischlag J.A.
      • Kyriakides T.C.
      • Padberg Jr., F.T.
      • Matsumura J.S.
      • Kohler T.R.
      • et al.
      Outcomes following endovascular vs open repair of abdominal aortic aneurysm: a randomized trial.
      The majority of re-interventions in our study were performed for iliac limb occlusion or stenosis. Van Keulen et al.
      • van Keulen J.W.
      • de Vries J.P.
      • Dekker H.
      • Goncalves F.B.
      • Moll F.L.
      • Verhagen H.J.
      • et al.
      One-year multicenter results of 100 abdominal aortic aneurysm patients treated with the Endurant stent graft.
      also found a higher iliac limb occlusion rate together with a 1-year re-intervention rate of 5% after implantation of an Endurant Stent Graft. At this moment, we are unable to conclude if there is any causality between occurrence of iliac occlusions and the wider inclusion criteria associated with the Endurant Stent Graft, or the stent graft itself.
      Despite numerous cases of short- or angulated necks, rates of early type-I endoleaks were low, with no ruptures and 42% of reported aneurysm-sac shrinkage. So far, few re-interventions were required to resolve endoleaks. A comparison with the literature on type-I/III re-interventions or shrinkage of the aneurysm awaits availability of longer-term follow-up.
      Compared with the landmark studies EVAR-1 and DREAM, in this study both perioperative mortality and the aneurysm-related mortality rates after 1 year were comparable. The all-cause mortality rate in this study compares favourably to EVAR-1 and DREAM.
      • EVAR-1 trial investigators
      Endovascular aneurysm repair versus open repair in patients with abdominal aortic aneurysm (EVAR trial 1): randomised controlled trial.
      • Prinssen M.
      • Verhoeven E.L.
      • Buth J.
      • Cuypers P.W.
      • van Sambeek M.R.
      • Balm R.
      • et al.
      A randomized trial comparing conventional and endovascular repair of abdominal aortic aneurysms.
      This is remarkable given the proportion of ASA class IV patients (10.6%) in the ENGAGE study population; in all previously assessed RCTs, ASA class IV patients were excluded.

      Conclusion

      ENGAGE is an unprecedented registry in scope and magnitude to characterise the performance of the Endurant Stent Graft in a contemporary series of 1262 AAA patients treated with the stent graft in 30 countries. The early results of the Endurant Stent Graft in a real-world, global experience are promising despite the fact that 10.6% of patients were classified ASA class IV and 17.9% of the patients were treated outside IFU. Longer follow-up is needed to assess durability of safety and effectiveness in a broader spectrum of AAA patients treated via endovascular methods.

      Registration

      ClinicalTrials.gov Identifier NCT00870051, registered since 25 March 2009.

      Contributions

      Study design: RS, JT, DB, VR, PH and MvS
      Data collection: JT, TF, DB, PP, VR, PH and MvS
      Data analysis: RS, JT, TF, VR and MvS
      Writing: RS, JT, TF, DB, PP, VR, PH and MvS

      Conflicts of Interest

      JT, TF, DB, PP, VR, PH and MvS have in the past received contributions from Medtronic AVE for giving oral presentations. JT, TF, DB, PP, VR and PH have been proctors for Medtronic AVE in the past.

      Financial Support

      The ENGAGE Registry was funded by Medtronic AVE. JT, DB, VR and MvS received an unrestricted research grant from Medtronic AVE.

      Appendix A.

      ENGAGE Participating investigators and centres. Argentina. Ricardo D. Garcia – Monaco, MD, PhD, Italian Hospital, Buenos Aires; Jaime Camacho, MD, Fundacion Favaloro, Buenos Aires. Australia. Rob Fitridge, MD, PhD, Queen Elizabeth Hospital, Adelaide/St Andrews Hospital, Adelaide; Michael J. Grigg, MD, PhD, Box Hill Hospital, Melbourne; Steve Dubenec, MD, Royal Prince Alfred Hospital, Camperdown; Nick Boyne, MD, Royal Brisbane and Womens Hospital, Brisbane; Michael J. Grigg, MD, PhD, Eastern Epworth Healthcare, Melbourne; Patrice B. Mwipatayi, MD, Royal Perth Hospital, Perth/Hollywood Private Hospital, Nedlands. Austria. Thomas Rand, MD, PhD, Hietzing Hospital, Vienna. Belgium. Patrick J. Peeters, MD, Imelda Hospital, Bonheiden; Jeroen M.H. Hendriks, MD, PhD, University Hospital, Antwerp; Marc Bosiers, MD, St. Blasius Hospital, Dendermonde; Frank Vermassen, MD, PhD, University Hospital, Ghent. Canada. Thomas L. Forbes, MD, London Health Sciences Centre Victoria Hospital, London; Min Lee, MD, Queen Elisabeth II Health Sciences Centre, Halifax; Leonard W.H. Tse, MD, Toronto General Hospital, Toronto; Yvan Douville, MD, Hospital Saint Francois d'Assise, Québec; Oren K. Steinmetz, MD, McGill University, Royal Victoria Hospital, Montréal. China. Jichun Zhao, MD, PhD, Hua Xi Hospital (West China Hospital), Sichuan; Jianfang Luo, MD, Guangdong General Hospital, Guangzhou; Wei Guo, MD, 301 PLA General Hospital, Bei Jing. Colombia. Jaime Camacho, MD, Fundacion Cardioinfantil, Bogota. Czech Republic. Jiri Novotny, MD, PhD, Institute for Clinical and Experimental Medicine, Prague. France. Jean-Pierre Becquemin, MD, Henri Mondor Hospital, Paris; Dominique Midy, MD, PhD, Pellegrin Hospital, Bordeaux; Emmanuel M. Choukroun, MD, Cardiology Hospital of Haut-Leveque, Pessac. Germany. Giovanni Torsello, MD, PhD, St Franziskus Hospital, Münster; Dittmar Böckler, MD, PhD, University Clinic, Heidelberg; Gerhard Hoffmann, MD, City Hospital, Solingen. Greece. Kostantinos O. Papazoglou, MD, PhD, Kianous Stavros Clinic, Thessaloniki; Dimitris A. Kiskinis, MD, Papageorgiou Hospital, Thessaloniki. Hong Kong. Stephen W.K. Cheng, MD, PhD, Queen Mary Hospital, Hong Kong. Israel. Yehuda G. Wolf, MD, Sourasky Medical Center, Tel Aviv. Italy. Andrea Stella, MD, S. Orsola – Malpighi Policlinic Hospital, Bologna; Carlo Setacci, MD, Le Scotte Policlinic Hospital, Siena; Carlo Pratesi, MD, Careggi Hospital, Florence. Korea. Do Yun Lee, MD, PhD, Yonsei Severence University Hospital, Seoul; Jae Kyu Kim, MD, PhD, Chonnam National University Hospital, Gwangju. Lithuania. Valdas Bilkis, MD, Vilnius University Hospital Santariskiu Klinikos, Vilnius. New Zealand. Andrew G. Hill, MD, Auckland City Hospital, Auckland; Thodur Vasudevan, MD, Waikato DHB Hospital, Hamilton. Norway. Eric J. Dorenberg, MD, Rikshospitalet University Hospital, Oslo; Rolf Busund, MD, PhD, University Hospital North-Norway, Troms; Guttorm L. Jenssen, MD, Haukeland University Hospital, Bergen. Poland. Walerian Staszkiewicz, MD, PhD, Bielanski Hospital, Warsaw; Piotr Gutowski, MD, PhD, Samodzielny Public Hospital Clinic 2, Poznan. Portugal. Joao Albuquerque e Castro, MD, Santa Marta Hospital, Lisbon. Slovakia. Ivan Vulev, MD PhD, Institute of Cardiovascular Diseases, Bratislava. South Africa. Jacobus Van Marle, MD, Unitas Hospital, Centurion; Corstiaan Leendert Punt, MD, St Georges Hospital, Port Elizabeth; Phillip J. Matley, MD, Kingsbury Hospital, Cape Town. Spain. Aurelio Garcia – de la Torre, MD, University Hospital Central – Asturias, Oviedo; Vincente Riambau, MD, PhD, University of Barcelona Hospital Clinic, Barcelona; Francisco J. Gomez – Palonés, MD, Peset University Hospital, Valencia; Carlos Vaquero – Puerta, MD, PhD, University Hospital, Valladolid; Eduardo Ros, PhD, San Cecilio University Hopspital, Granada. Sweden. HåkanRoos, MD, Sahlgrenska University Hospital, Gothenburg; Martin Delle, MD, PhD, Sodersjukhuset, Stockholm; Thomas Larzon, MD, Orebro University Hospital, Orebro. Switzerland. Do Dai-Do, MD, PhD, Inselspital, University Hospital, Berne. Thailand. Boonprasit Kritpracha, MD, Songklanagarind Hospital, Songkhla. The Netherlands. Joep A.W. Teijink, MD, PhD, Catharina Hospital, Eindhoven; Hence J.M. Verhagen, MD, PhD, Erasmus Medical Center, Rotterdam; Steven M.M. van Sterkenburg, MD, Rijnstate Hospital, Arnhem; Rob Welten, MD, PhD, Atrium Medical Center, Heerlen; Jan M.M. Heijligers, MD, PhD, St. Elisabeth Hospital, Tilburg; Jean-Paul P.M. de Vries, MD, PhD, St. Antonius Hospital, Nieuwegein. Turkey. A. Yigit Göktay, MD, Dokuz Eylül University Hospital, Izmir; Levent Oguzkurt, MD, Adana Baskent University Hospital, Adana; Furuzan Numan, MD, PhD, Memorial Hospital, Istanbul/Istanbul University Cerrahpasa, Istanbul. United Kingdom. Paul D. Hayes, MD, Addenbrooke’s Hospital, Cambridge; John Rose, MD, Freeman Hospital, Newcastle upon Tyne; Richard G. McWilliams, MD, Royal Liverpool University Hospital, Liverpool; Ray Ashleigh, MD, Wythenshawe Hospital, Manchester; Matt Thompson, MD, PhD, St George’s Hospital Tooting, London. Uruguay. Carmelo Gastambide, MD, Medical Corporation Uruguaya, Montevideo.

      References

        • Lederle F.A.
        • Johnson G.R.
        • Wilson S.E.
        • Chute E.P.
        • Littooy F.N.
        • Bandyk D.
        • et al.
        Prevalence and associations of abdominal aortic aneurysm detected through screening. Aneurysm Detection and Management (ADAM) Veterans Affairs Cooperative Study Group.
        Ann Intern Med. 1997; 126: 441-449
        • Parodi J.C.
        • Palmaz J.C.
        • Barone H.D.
        Transfemoral intraluminal graft implantation for abdominal aortic aneurysms.
        Ann Vasc Surg. 1991; 5: 491-499
        • Volodos N.L.
        • Karpovich I.P.
        • Troyan V.I.
        • Kalashnikova Yu V.
        • Shekhanin V.E.
        • Ternyuk N.E.
        • et al.
        Clinical experience of the use of self-fixing synthetic prostheses for remote endoprosthetics of the thoracic and the abdominal aorta and iliac arteries through the femoral artery and as intraoperative endoprosthesis for aorta reconstruction.
        Vasa Suppl. 1991; 33: 93-95
        • Chaikof E.L.
        • Brewster D.C.
        • Dalman R.L.
        • Makaroun M.S.
        • Illig K.A.
        • Sicard G.A.
        • et al.
        The care of patients with an abdominal aortic aneurysm: the Society for Vascular Surgery practice guidelines.
        J Vasc Surg. 2009; 50: S2-49
        • EVAR-1 trial investigators
        Endovascular aneurysm repair versus open repair in patients with abdominal aortic aneurysm (EVAR trial 1): randomised controlled trial.
        Lancet. 2005; 365: 2179-2186
        • Prinssen M.
        • Verhoeven E.L.
        • Buth J.
        • Cuypers P.W.
        • van Sambeek M.R.
        • Balm R.
        • et al.
        A randomized trial comparing conventional and endovascular repair of abdominal aortic aneurysms.
        N Engl J Med. 2004; 351: 1607-1618
        • Lederle F.A.
        • Freischlag J.A.
        • Kyriakides T.C.
        • Padberg Jr., F.T.
        • Matsumura J.S.
        • Kohler T.R.
        • et al.
        Outcomes following endovascular vs open repair of abdominal aortic aneurysm: a randomized trial.
        JAMA. 2009; 302: 1535-1542
        • Becquemin J.P.
        • Pillet J.C.
        • Lescalie F.
        • Sapoval M.
        • Goueffic Y.
        • Lermusiaux P.
        • et al.
        A randomized controlled trial of endovascular aneurysm repair versus open surgery for abdominal aortic aneurysms in low- to moderate-risk patients.
        J Vasc Surg. 2011; 53 (1167–1173.e1)
        • Choke E.
        • Munneke G.
        • Morgan R.
        • Belli A.M.
        • Loftus I.
        • McFarland R.
        • et al.
        Outcomes of endovascular abdominal aortic aneurysm repair in patients with hostile neck anatomy.
        Cardiovasc Intervent Radiol. 2006; 29: 975-980
        • Carpenter J.P.
        • Baum R.A.
        • Barker C.F.
        • Golden M.A.
        • Mitchell M.E.
        • Velazquez O.C.
        • et al.
        Impact of exclusion criteria on patient selection for endovascular abdominal aortic aneurysm repair.
        J Vasc Surg. 2001; 34: 1050-1054
        • Rouwet E.V.
        • Torsello G.
        • de Vries J.P.
        • Cuypers P.
        • van Herwaarden J.A.
        • Eckstein H.H.
        • et al.
        Final results of the prospective European trial of the Endurant stent graft for endovascular abdominal aortic aneurysm repair.
        Eur J Vasc Endovasc Surg. 2011; 42: 489-497
        • Bockler D.
        • Fitridge R.
        • Wolf Y.
        • Hayes P.
        • Silveira P.G.
        • Numan F.
        • et al.
        Rationale and design of the Endurant Stent Graft Natural Selection Global Postmarket Registry (ENGAGE): interim analysis at 30 days of the first 180 patients enrolled.
        J Cardiovasc Surg (Torino). 2010; 51: 481-491
        • van Marrewijk C.J.
        • Leurs L.J.
        • Vallabhaneni S.R.
        • Harris P.L.
        • Buth J.
        • Laheij R.J.
        Risk-adjusted outcome analysis of endovascular abdominal aortic aneurysm repair in a large population: how do stent-grafts compare?.
        J Endovasc Ther. 2005; 12: 417-429
        • Ligthelm R.J.
        • Borzi V.
        • Gumprecht J.
        • Kawamori R.
        • Wenying Y.
        • Valensi P.
        Importance of observational studies in clinical practice.
        Clin Ther. 2007; (29 Spec No): 1284-1292
        • Kvinlaug K.E.
        • Lawlor D.K.
        • Forbes T.L.
        • Willoughby R.
        • MacKenzie K.S.
        • DeRose G.
        • et al.
        Early results from a Canadian multicenter prospective registry of the Endurant stent graft for endovascular treatment of abdominal aortic aneurysms.
        J Endovasc Ther. 2012; 19: 58-66
        • Torsello G.
        • Troisi N.
        • Tessarek J.
        • Torsello G.F.
        • Dorigo W.
        • Pulli R.
        • et al.
        Endovascular aortic aneurysm repair with the Endurant stent-graft: early and 1-year results from a European multicenter experience.
        J Vasc Interv Radiol. 2010; 21: 73-80
        • van Keulen J.W.
        • de Vries J.P.
        • Dekker H.
        • Goncalves F.B.
        • Moll F.L.
        • Verhagen H.J.
        • et al.
        One-year multicenter results of 100 abdominal aortic aneurysm patients treated with the Endurant stent graft.
        J Vasc Surg. 2011; 54: 609-615
        • De Bruin J.L.
        • Baas A.F.
        • Buth J.
        • Prinssen M.
        • Verhoeven E.L.
        • Cuypers P.W.
        • et al.
        Long-term outcome of open or endovascular repair of abdominal aortic aneurysm.
        N Engl J Med. 2010; 362: 1881-1889
        • Peterson B.G.
        • Matsumura J.S.
        • Brewster D.C.
        • Makaroun M.S.
        Five-year report of a multicenter controlled clinical trial of open versus endovascular treatment of abdominal aortic aneurysms.
        J Vasc Surg. 2007; 45: 885-890
        • Prinssen M.
        • Buskens E.
        • de Jong S.E.
        • Buth J.
        • Mackaay A.J.
        • van Sambeek M.R.
        • et al.
        Cost-effectiveness of conventional and endovascular repair of abdominal aortic aneurysms: results of a randomized trial.
        J Vasc Surg. 2007; 46: 883-890

      Linked Article

      Comments

      Commenting Guidelines

      To submit a comment for a journal article, please use the space above and note the following:

      • We will review submitted comments as soon as possible, striving for within two business days.
      • This forum is intended for constructive dialogue. Comments that are commercial or promotional in nature, pertain to specific medical cases, are not relevant to the article for which they have been submitted, or are otherwise inappropriate will not be posted.
      • We require that commenters identify themselves with names and affiliations.
      • Comments must be in compliance with our Terms & Conditions.
      • Comments are not peer-reviewed.