If you don't remember your password, you can reset it by entering your email address and clicking the Reset Password button. You will then receive an email that contains a secure link for resetting your password
If the address matches a valid account an email will be sent to __email__ with instructions for resetting your password
Department of Vascular and Endovascular Surgery, Royal Adelaide Hospital, Port Road, Adelaide, SA 5000, AustraliaThe University of Adelaide, SA, Australia
The aim was to determine whether lead containing and lead free composite garments in current use provide the level of radiation protection stated by manufacturers.
Methods
Fifteen garments, produced by five different manufacturers using eight different composites, were randomly selected for testing from four hospitals in South Australia. Labelling, material composition, design, and condition of the garments were assessed by direct garment examination, garment label, and product information. Garment attenuation was tested in a simulated angiography suite using a Siemens Ysio Max digital Xray machine. The front and back panels of each garment were tested under direct beam at 100 kVp. A Perspex phantom was used to simulate the density and scatter properties of the human abdomen. The front panels of each garment were tested under scattered radiation at Xray tube voltages of 50 and 70 kVp.
Results
Forty-seven per cent of front panels and 90% of back panels provided lower lead equivalence than claimed by the manufacturer. Twenty per cent of front panels and 62% of back panels tested did not meet the minimum International Electrotechnical Commission requirements for angiographic use. There was a 38 fold difference in front panel performance of garments to scatter radiation, which were all labelled 0.5 mm lead equivalence. 56% of garments had differences in scatter transmission of at least 49% when tested at 50 and 70 kVp.
Conclusion
The results show that lead containing and lead free composite garments probably provide less radiation protection than manufacturer stated lead equivalence. The demonstrated wide variations in attenuation of scatter radiation are greater than previously reported. It was found that most garments failed to comply with labelling standards. The study highlights challenges in radiation shielding and the need to identify composites that consistently provide better attenuation per unit weight than lead.
The aim of this study was to test the performance and labelling of garments that are in contemporary use for radiation protection. A significant proportion of garments failed to comply with international labelling standards and did not provide the protection claimed by manufacturers. In addition, garments labelled at the same lead equivalence performed vastly differently in attenuation of scatter. These findings were demonstrated in the real world setting, by testing garments that are in current use across multiple centres. The study highlights current concerns regarding radiation protection standards and the need to identify composites that can consistently provide better protection than lead.
Introduction
The number and complexity of procedures performed using interventional radiology is increasing.
These procedures can be prolonged and require staff to be in close proximity to the patient. In the working environment, the interventionist should, under no circumstances, be subject to direct beam exposure. Provided basic radiation protection principles are followed, radiation exposure to staff is likely to be entirely due to the scattered beam. Adequate radiation protection for healthcare workers, especially those performing interventional radiology procedures, is critical. Lead has traditionally been used in protective garments and shields across the range of radiation energies typically found in diagnostic and interventional procedures.
Lead is particularly suitable for radiation protection because of its high absorption and attenuation of Xray photons and its relatively cheap cost to manufacture. However, lead is a heavy material, and when worn can lead to chronic musculoskeletal disorders.
These musculoskeletal effects may reduce the quality of life and productivity of workers.
Given the problems associated with lead's weight, there is ongoing interest in developing lighter weight alternatives, including lead containing and lead free composites. Combinations of metals including antimony, bismuth, tin, aluminium, tungsten, titanium, and barium have been used. Lead only garments have largely been surpassed in favour of lighter weight alternatives that offer 25–40% weight reductions and are more flexible.
However, verifying manufacturer claims about how these garments compare to traditional lead garments regarding their protective properties can be challenging. In addition, the performance of composites has been mixed, with composites often underperforming relative to the stated lead equivalence.
Lead ambivalence of radiation protection apparel: when is bad good? Combined scientific meeting imaging and radiation in personalised medicine, 1 september 2014; Melbourne, Australia.
The level of protection provided by protective garments is described in terms of millimetres of lead equivalence (mm Pb LE), defined as the “thickness of lead, which would attenuate the same amount of radiation as the given material when exposed to radiation of the same type and quality”.
The attenuation coefficient of a material determines how easily it can be penetrated by radiation, and is influenced by structural forces at atomic level, which, in turn, are highly dependent on the tube voltage that the material is tested at.
Most national standards, including European standards, are based on the International Electrotechnical Commission (IEC) International Standard 61331–3:2014.
The standard states that the thickness of “light workload” garments should be at least 0.25 mm Pb LE over the entire area and for “heavy workload” garments it should be at least 0.35 mm Pb LE at the front. Heavy workload use refers to angiographic type work and light workload encompasses all other work.
The primary aim of this study was to determine whether composite garments in current use provide the level of protection stated by manufacturers.
Materials and Methods
The study aim was addressed by assessing garment (i) labelling; (ii) material composition; (iii) design; (iv) condition; (v) direct beam attenuation; and (vi) scatter attenuation. Owing to the heterogenicity of the published data, a meaningful power calculation was not possible. Fifteen garments were chosen as the maximum number available for testing without disrupting the daily workflow of the departments from which they were sourced. These were randomly selected for testing from four public hospitals within Adelaide, South Australia. They were a representative sample of garments used in angiography suites at the four sites. They were all transported and tested at the same facility under the same conditions on one day.
Labelling
The labelling of garments was tested against IEC International Standard 61331–3:2014.
The standard states that garments should be labelled with (i) name/trademark of the manufacturer or supplier; (ii) area density; (iii) attenuation equivalent; (iv) Xray tube voltage (and filtration); (v) size of garment; (vi) reference to the standard; and (vii) workload (heavy or light).
Material composition
The garment manufacturer and the proportion of garments where the exact material was stated on the label was recorded. If the material of the gown was not stated, an attempt was made to identify it using the manufacture's product information.
Design
A description of each garment was made and morphometric assessment of garment design regarding efficacy of coverage of the body was performed.
Condition
Owing to the potential confounding effect of a garment's physical condition on the results, an assessment was made by three investigators regarding the overall condition based on subjective degree of “wear and tear”: (i) heavy; (ii) medium; (iii) light; (iv) new.
Table 1 Provides a description of the garments examined.
Table 1Description of garments used in 15 investigated apron models, including composition of garment, compliance with International Electrotechnical Commission (IEC) Standards and lead equivalence values
Garment attenuation was tested in a simulated angiography suite using a Siemens Ysio Max (Siemens Healthcare, Erlangen, Germany) digital Xray machine with field area of 10 cm × 10 cm. Garments were placed 100 cm from the Xray machine and a Raysafe Xi solid state radiation detector (Unfors RaySafe AB, Billdal, Sweden) was placed behind the garment prior to exposure to the direct beam (Fig. 1). Prior to testing, repeated measures of each parameter were performed to confirm equipment performance. Consistency was excellent with negligible differences; coefficients of variation were 0.23% kVp, 0.06% exposure time, and 0.1% tube output.
Figure 1Set up for scatter attenuation measurement (A) and direct beam attenuation measurement (B).
The front and back panels of each garment were tested under direct beam at 100 kVp. The duration of each Xray was 100 ms and the resulting measurement was an average of three separate measurements. The lead equivalence of each garment was calculated from its attenuation with no apron present, and by repeating this with high purity lead foils, the lead equivalence could be determined using the Archer equation.
The use of broad beam geometry in the measurement incorporates the production of scatter within the material being investigated. In long garments, measurements were taken at the chest to measure dose to the radiosensitive organs of interest. In garments with an overlapping front section, measurements were taken at the overlap section as this is the part which the manufacturer stated lead equivalence refers to.
Scatter attenuation
A Perspex phantom (17.5 cm × 17.5 cm × 29 cm) was used to simulate the density and scatter properties of the human abdomen. Garments were placed below and 50 cm away from the phantom to simulate the position of an operator in relation to a patient on an angio table. A Raysafe XI radiation detector was placed inside the garment to measure the amount of scattered radiation reaching the operator (Fig. 1). In order to compare a homogenous group of garments, scattered radiation attenuation was tested by front panels only of the 9 garments labelled 0.5 mm Pb LE. Xray tube voltages of 50 kVp and 70 kVp were used, with a 200 ms exposure duration. Results were reported with a transmission coefficient, given as a percentage.
Results
Labelling
Compliance to IEC International Standard 61331–3:2014 labelling standard is detailed in Table 2. The majority (80%) of labels failed the standard. Eighty per cent failed to state whether the garment was intended for heavy or light use. The attenuation equivalence and Xray tube testing voltage were each missing in 7%. Thirty-three per cent did not have sizing labels.
Table 2Compliance with International Electrotechnical Commission (IEC) International Standard 61331–3:2014 by labelling item
Most vests and skirts have an area of overlap at the front where the material thickness is doubled. Many manufactures rely on this overlapping design to provide the required minimum thickness for the front panel.
Labels were unclear regarding whether the stated lead equivalence referred to the single layer or overlapping area. This apparently simple overlap design varied significantly among garments tested and was fundamental to whether the vest provided adequate coverage. If the overlap was large then the degree of protection was adequate; however, a narrow overlap led to a large area of the vest failing to provide adequate protection, despite purported front panel protection of 0.5 mm Pb (Fig. 2).
Figure 2Vest with a large area of overlap resulting in minimal areas of reduced protection at the side (A) vs. vest with small area of overlap resulting in large areas of reduced protection at the side (B). Pb = lead.
The 15 garments were produced by five different manufacturers using eight different composites. These composites were identified by proprietary names and included lead containing and lead free composites. None of the labels provided information regarding material composition. Details from manufacturers pertaining to only three composites were found. Descriptions included containing “antimony and bismuth” and containing “70–90% lead and 10–30% plastisol”.
Design
The 15 garments consisted of five different designs: open back apron (n = 2), closed back apron (n = 3), poncho (n = 1), vest (n = 6), and skirt (n = 3).
Direct beam attenuation
Front panel
All 15 garments had front panels and were tested. Seven garments (47%) provided lower than the stated attenuation equivalence (Fig. 3). Three (20%) garments provided < 0.35 mm Pb LE protection and fell below the minimum IEC requirement for front panel angiography use.
Figure 3Front panel lead equivalence to direct beam. Each garment has been assigned an arbitrary identification number (1–15) on the x axis and the manufacturer's stated protection (mm Pb LE) is displayed at the bottom (solid box). The actual measured protection (mm Pb LE) is displayed directly beneath (dotted box). The bar graphs represent the percentage difference of the actual protection compared with the manufacture's stated. Garments that performed better than expected are depicted as a bar above the x-axis in white, and those that performed worse as a bar below the x axis in red. Garment no. 3 measured an unrecordable attenuation and therefore performed better than stated to an undetermined degree. Garment no.14 had no manufacture's stated protection making comparison impossible but performed well at a measured protection of 0.6 mm Pb LE. Pb = lead.
Two garments did not have a back panel and therefore were not tested. Three garments with a back panel did not display a stated lead equivalence and therefore comparisons with the manufacturer's stated protection could not be made. When testing the remaining 10 garments with the manufacturer's stated back panel equivalence, nine (90%) provided lower than the stated attenuation equivalence (Fig. 4). Of the 13 garments with back panels, eight (62%) provided < 0.25 mm Pb protection and fell below the minimum IEC requirements for back panels.
Figure 4Back panel lead equivalence to direct beam. Each garment has been assigned an arbitrary identification number (1–15) on the x axis and the manufacturer's stated protection (LE, Pb mm) is displayed at the bottom (solid box). The actual measured protection (LE, Pb mm) is displayed directly beneath (dotted box). The bar graphs represent the percentage difference of the actual protection compared with the manufacturer's stated. Garments that performed better than expected are depicted as a bar above the x axis in white, and those that performed worse as a bar below the x axis in red. Garment nos: 1 and 9 had no back and were therefore not tested. Garment nos: 2, 7, and 14 did not display any stated manufacturer's protection for the back panel and therefore a comparison could not be made. Pb = lead.
Attenuation of scattered radiation was highly variable among the nine garments tested (Fig. 5). Transmission varied up to 38 fold from 0.05% to 1.90% between garments with purported identical protection according to the manufacturer. Individual garments performed significantly differently and unpredictably when tested at the two different Xray tube voltages, 50 kVp and 70 kVp. Five of the nine garments (56.6%) had differences in transmission of at least 49% between these two energies. For example, garment no. 6 permitted a 52% increase in transmission at a lower testing energy (0.79% at 50 kVp vs. 0.52% at 70 kVp), whereas garment no. 7 permitted a 388% increase in transmission at the higher energy (0.49% transmission at 50 kVp vs. 1.90% transmission at 70 kVp).
Figure 5Transmission of scattered radiation at tube voltages 50 kVp and 70 kVp by front panels of 9 garments labelled 0.5 LE mmPb. Pb = lead.
The condition of the garments was well distributed between heavy (n = 4), medium (n = 4), light (n = 2), and new (n = 5) in appearance. All gowns are assigned a 10 year “lifespan” prior to decommission and all aprons had been purchased within this time period. Although there was a trend towards decreased attenuation in front panel with older garments (p = .32), the converse was true for the back panel (p = .004). Owing to the small numbers involved, additional statistical analysis was made comparing an older group (“heavy” and “medium”) to a “newer” group (“light” and “new”). Applying the Mann–Whitney and Krukal–Wallis tests, there was no statistically significant difference in attenuation with subjective assessment of garment condition (Table 3).
Table 3Effect of garment condition on transmission attenuation (measured in Pb LE)
Radiation physics and performance of composite garments
The attenuation coefficient is used to describe how strongly a material interacts with radiation. Generally, the attenuation coefficient decreases with lower density materials and with higher incident photon energy. The “K-edge” is the binding energy of the innermost electron shell of the atoms within the material being tested. At this energy level, there is an abrupt increase in attenuation coefficient such that the photons just above this energy level are much more easily absorbed than photons just below this energy level. This non-linear behaviour is depicted as “spikes” on attenuation coefficient figures. The K-edge is highly variable among different materials and this has a significant influence on the attenuation coefficient, especially when exposed to the range of photon energies used in medical radiology. Xray tube voltages, which determine the incident photon energy, typically range from 60 to 125 kVp for diagnostic radiology and up to 140 kVp for angiography.
The attenuation coefficient of lead has a relatively linear relationship with incident photon energy levels within the range of tube voltages typically used, with little K-edge influence (Fig. 6). Other elements used in composites, such as antimony, do not have such a predictable and linear behaviour, and the K-edge of the material becomes highly influential. In order to compensate for this phenomenon, manufacturers use multiple materials with different attenuation coefficients to provide an “overlap” of protection to exploit the K-edge.
Figure 6Relationship between attenuation coefficient (μ/ρ in cm
/g) and photon energy (MeV) for lead (left) and antimony (right). The white area highlights the typical range of energy levels encountered in diagnostic radiology. Adapted from X-Ray Mass Attenuation Coefficients, National Institute of Standards and Technology
Lead ambivalence of radiation protection apparel: when is bad good? Combined scientific meeting imaging and radiation in personalised medicine, 1 september 2014; Melbourne, Australia.
Although some studies have found that composites are equivalent or better than lead at certain tube voltages it must be stressed that for adequate protection garments need to provide protection across a wide range of Xray voltages.
This study has demonstrated a wide variation in compliance to labelling standards and protection provided by radiation protection garments tested in a real world setting.
It was found that most garments failed to comply with labelling standards. Given labelling requirements in most countries, including Australia, are based on IEC standards, it is possible poor labelling compliance is present worldwide.
Overall, the tested garments generally performed poorly both in terms of achieving the lead equivalence claimed by manufacturers, as well as meeting the minimum IEC requirements. Front panels generally provided much better protection than back panels, with 53% outperforming manufacturer stated lead equivalence. However, 20% still did not meet the minimum IEC requirements for angiographic use. Back panels performed poorly, with 90% underperforming according to manufacturer's claims, and 62% not meeting the minimum IEC requirements for angiographic use.
One criticism of current garment testing protocols is the heavy reliance on narrow direct beam testing. What is of much more concern is exposure to the more unpredictable scattered radiation, which is rarely tested. This study highlighted the concerning finding of a 38 fold difference in performance of garments to scatter, which purportedly all provided the same level of protection. This wide discrepancy is much larger than previously documented.
Impact of using scatter-mimicking beams instead of standard beams to measure penetration when assessing the protective value of radiation-protective garments.
Furthermore, protection against scattered radiation was shown to be highly variable and unpredictable at different energies. While it is impossible to determine the absolute health risk that these findings represent, simple calculations assuming the linear no threshold model of radiation biology would suggest that it is statistically possible a fatal cancer could be induced by the widespread use of substandard radiation protection apparel. Therefore, it is important to apply vigilantly ALARA radiation protection principles (“as low as reasonability achievable”) in the workplace in conjunction with providing appropriate protective garments and other forms of shielding to staff. These include a reduction of fluoroscopy and angiography duration, avoiding steep angulation, increasing the distance between healthcare workers and the radiation source, reducing magnification and frame rate, and the use of collimation.
Limitations
This study tested garments currently used in hospitals in a “real world” setting, which inevitably led to several limitations. Although an attempt was made to choose a selection of different garment designs and manufacturers, the choice was inevitably random and subject to availability. However, a good selection of garments has been demonstrated, varying in condition from heavily worn to new. Although the exact age of the garments within their 10 year procurement could not be confirmed, subgroup analysis failed to show that condition significantly affected the results (Table 3). Although shielding integrity looking for cracks in the garments was not performed in this study, as part of accreditation requirements, the relevant national codes of practice dictate that aprons within each health organisation are screened for shielding integrity annually and faulty garments removed.
The testing environment was a simulated angiography suite. A real angiography suite could not be used as the Xray beam energy is automatically set by the machine, preventing measurement consistency and comparison of results. Therefore general radiographic apparatus was used with manual control of beam filtration, tube current, and photon energies. Although the Xray source was above the phantom (opposite to a real laboratory), to compensate for this discrepancy the garment and detector were positioned below the phantom in the equivalent position that an operator would be standing. While the use of a simulated laboratory failed to test the performance of the gowns in daily practice, it allowed for control of the numerous confounding variables that influence radiation exposure, resulting in an experimental set up focused on performance of gowns based on composition, design, and condition only.
It was hoped that testing could be performed across the range of Xray energies typically used in interventional radiology (60–140 kVp). This was not possible owing to a significant range of radiation energy and intensity between primary beam, scattered radiation, and attenuated scattered radiation. Although the use of multiple dosimeters would allow measurements across all energy and intensity ranges, comparison would be difficult. Although devices with dose and energy response ranges of this magnitude exist, such as large volume ionisation chambers, they have limited utility in a hospital environment and a prohibitively high cost. Therefore, the study was limited to the ranges measured, although the large variations in protection afforded even with these limited energies, was adequately shown.
The impact of scatter can affect direct beam attenuation measurement. In order to minimise this, the experiment was set up so that the location of the measurement probe would be minimally impacted by scatter reaching it over or around the apron. In addition the use of broad beam geometry in the measurement, incorporated the production of scatter radiation within the material being investigated.
Finally, the exact composition of all eight of the composites tested could not be determined, and the study only determined nebulous details of three. This uncertainty does not alter the overall findings, and, in fact, highlights the difficulty in determining exactly what the garments are made of and how much protection they actually provide.
Conclusions
Radiation exposure can be significantly reduced by wearing protective garments, utilising mounted shields,
Operator radiation and the efficacy of ceiling-suspended lead screen shielding during coronary angiography: an anthropomorphic phantom study using real-time dosimeters.
and adhering to ALARA principals. Currently, however, the complex and ambiguous nature of “lead equivalence” as a descriptor of radiation protection may result in confusion when choosing protective garments. It is recommended that garment labels better reflect the level of protection provided, and should include lead equivalence and transmission values at a range of photon energies at which the garment is designed to be used, including overlap and non-overlap segments. Until then it is possible that garments will continue to fail to provide the protection that users expect and deserve.
Conflict of interest
None.
Funding
None.
References
The Royal College of Radiologists and British Society of Interventional Radiologists
Investing in the Interventional Radiology Workforce: the Quality and Efficiency Case London.
Lead ambivalence of radiation protection apparel: when is bad good? Combined scientific meeting imaging and radiation in personalised medicine, 1 september 2014; Melbourne, Australia.
Impact of using scatter-mimicking beams instead of standard beams to measure penetration when assessing the protective value of radiation-protective garments.
Operator radiation and the efficacy of ceiling-suspended lead screen shielding during coronary angiography: an anthropomorphic phantom study using real-time dosimeters.
Endovascular procedures have drastically increased in number and complexity during the past decade. While the pioneers of endovascular repairs probably knew about the risks associated with ionizing radiation exposure, the initial enthusiasm somehow concealed the danger and set aside physician safety. While the benefit/risk ratio regarding the exposure to ionizing radiation is often obvious for our patients, we have to keep in mind that the risk is completely unbalanced when it comes to the safety of physicians and medical staff.
To submit a comment for a journal article, please use the space above and note the following:
We will review submitted comments as soon as possible, striving for within two business days.
This forum is intended for constructive dialogue. Comments that are commercial or promotional in nature, pertain to specific medical cases, are not relevant to the article for which they have been submitted, or are otherwise inappropriate will not be posted.
We require that commenters identify themselves with names and affiliations.